Introduction

Practitioners use the terms sudden cardiac arrest (SCA) and sudden cardiac death interchangeably; however, their definitions are distinct. Sudden cardiac arrest is the “sudden cessation of cardiac activity so that the victim becomes unresponsive, with no normal breathing and no signs of circulation.” If the condition is not addressed immediately, sudden cardiac arrest progresses to sudden cardiac death. Sudden cardiac death is defined “as a natural death due to cardiac causes, heralded by abrupt loss of consciousness.” Out-of-hospital cardiac arrest occurs outside of the hospital, and emergency medical services personnel most commonly attend to patients in these situations. There are approximately 356,500 out-of-hospital cardiac arrests per year in the United States.[1][2][3]

Anatomy and Physiology

Register For Free And Read The Full Article
Get the answers you need instantly with the StatPearls Clinical Decision Support tool. StatPearls spent the last decade developing the largest and most updated Point-of Care resource ever developed. Earn CME/CE by searching and reading articles.
  • Dropdown arrow Search engine and full access to all medical articles
  • Dropdown arrow 10 free questions in your specialty
  • Dropdown arrow Free CME/CE Activities
  • Dropdown arrow Free daily question in your email
  • Dropdown arrow Save favorite articles to your dashboard
  • Dropdown arrow Emails offering discounts

Learn more about a Subscription to StatPearls Point-of-Care

Anatomy and Physiology

Confirmed by a series of trials (Antiarrhythmics versus Implantable Defibrillators (AVID), Cardiac Arrest Study Hamburg (CASH), and the Canadian Implantable Defibrillator Study (CIDS)), implantable cardioverter-defibrillators or ICDs have been proven significantly to reduce the risk of sudden cardiac death attributable to:

  • In patients with New York Heart Association (NYHA) class I, II, or III Ventricular tachyarrhythmias caused by a prior myocardial infarction
  • In patients with New York Heart Association class II or III ventricular tachyarrhythmias caused by non-ischemic cardiomyopathy despite optimal medical therapy, and in whom survival with good functional capacity is anticipated to extend beyond 1 year.

The ICD improved survival despite beta-blockade, surgical revascularization, or presenting arrhythmia (VT or VF). There was no difference in benefit gained from ICD implantation between those patients with coronary disease and those with non-ischemic cardiomyopathies. Primary prevention trials also played an important role in proving the unique benefit of ICD implantation even before the incidence of lethal arrhythmia, patients with left-ventricle (LV) dysfunction and post-myocardial infarction and a history of non-sustained VT underwent electrophysiology (EP) testing to identify higher risk patients with inducible, non-suppressible, ventricular tachyarrhythmias. This population was the target for the primary prevention trials, including:

  • The famous Multicenter Automatic Defibrillator Trial (MADIT) compared ICDs to conventional therapy (mainly amiodarone) (ejection fraction (EF) =35%)
  • The Multicenter Unsustained Tachycardia Trial (MUSTT) compared EP-guided therapy (ICDs or drug therapy) to no EP-guided therapy (EF =40%).

The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) confirmed the benefit of ICDs in ischemic patients. This validated findings in MADIT II and those of a previous smaller study of non-ischemic cardiomyopathy patients, the Prophylactic Defibrillator Implantation in Patients with Non-ischemic Dilated Cardiomyopathy (DEFINITE) trial.

Doctors use ICDs to prevent sudden cardiac death in other high-risk patient subsets, such as patients with ion-channel abnormalities such as Brugada syndrome, long QT syndrome (LQTS), or other structural heart diseases such as RV dysplasia and hypertrophic cardiomyopathy. Although prospective randomized trials in these rare conditions are not likely to be pursued, case series showed the efficacy of the ICD in patients with LQTS, Brugada syndrome, hypertrophic cardiomyopathy, and arrhythmogenic RV dysplasia.[4][5][6]

Indications

Indications of ICD according to American College of Cardiology/American Heart Rhythm Society (ACC/AHRS) 2008:[7][8][9][10]

Class I in patients having any of the following:

  1. Survivors of cardiac arrest due to hemodynamically unstable sustained ventricular tachycardia (VT) or FF after evaluation and exclusion of any reversible cause
  2. Structural heart disease with spontaneous sustained VT, whether hemodynamically stable or not
  3. Unexplained syncope, with hemodynamically significant sustained VT or ventricular fibrillation (VF), induced at electrophysiology study (EPS)
  4. Ischemic cardiomyopathy patients with left-ventricular ejection fraction (LVEF) less than 35% due to prior MI who are a minimum of 40 days post-MI and are in NYHA functional class II or III.
  5. Non-ischemic dilated cardiomyopathy (DCM) with LVEF 35% or less and in NYHA functional class II or III.
  6. Patients with LV dysfunction due to prior MI who are a minimum of 40 days post-MI and LVEF less than 30%, with New York Heart Association functional class I
  7. Non-sustained VT due to prior MI, LVEF less than 40%, and sustained VT or inducible VF at EPS.

Class IIa

  1. Unexplained syncope, non-ischemic DCM, and significant LV dysfunction
  2. Patients with sustained VT and normal or near-normal ventricular function
  3. Hypertrophic cardiomyopathy who have 1 or more major risk factors for sudden cardiac death
  4. Prevention of sudden cardiac death in patients with arrhythmogenic right ventricular dysplasia (ARVD/C) who have 1 or more risk factors for sudden cardiac death
  5. Patients with Long QT syndrome, to reduce sudden cardiac death, who are suffering syncope or Ventricular tachycardia
  6. Nonhospitalized patients are awaiting transplantation
  7. Brugada syndrome patients who have had syncope or have documented VT that has not resulted in cardiac arrest
  8. Catecholaminergic polymorphic VT who have syncope or documented sustained VT while receiving beta-blockers
  9. Patients with cardiac sarcoidosis, Chagas disease, and giant cell myocarditis.

Class IIb

  1. Non-ischemic heart disease with LVEF 35% and who are in NYHA functional class I.
  2. Long-QT syndrome patients and risk factors for sudden cardiac death
  3. ICD therapy may be considered in patients with sudden cardiac death and advanced structural heart disease who, without a defining cause by invasive or noninvasive methods
  4. Patients with a familial cardiomyopathy associated with sudden death
  5. Patients with LV noncompaction.

Contraindications

Contraindications include the following:

  1. ICD therapy is not indicated in those who do not have a reasonable expectation of survival with an acceptable functional status for at least 1 year, even if they meet implantation criteria for ICD specified in class I, IIa, and IIb recommendations above. (Level of Evidence: C)
  2. ICD therapy is not recommended for patients with incessant VT or VF. (Level of Evidence: C)
  3. ICD therapy is not recommended in significant psychiatric illnesses that may be aggravated by device implantation or that may preclude systematic follow-up. (Level of Evidence: C)
  4. ICD therapy is not recommended in NYHA Class IV with drug-refractory congestive heart failure who are not candidates for cardiac transplantation or CRT-D. (Level of Evidence: C)
  5. ICD therapy is not recommended for syncope of undetermined cause in a patient without inducible ventricular tachyarrhythmias and structural heart disease. (Level of Evidence: C)
  6. ICD therapy is not recommended when VF or VT is amenable to surgical or catheter ablation (eg, atrial arrhythmias associated with the Wolff-Parkinson-White syndrome, LV, or RV RV outflow tract VT, idiopathic VT, or fascicular VT in the absence of structural heart disease). (Level of Evidence: C)
  7. ICD therapy is not recommended for patients with ventricular tachyarrhythmias as a result of an entirely reversible disorder in the absence of structural heart disease (eg, drugs, electrolyte imbalance, or trauma).

Equipment

ICD Components

The ICD generator

The ICD generator has the necessary elements to coordinate all pacing, sensing, and defibrillation functions. These elements include the battery and electronic circuitry (brains and capacitors).

  • The Battery: Each ICD has a battery whose lifetime depends on how much pacing and defibrillating the device is called upon to perform. Typical longevities fall between about 4 to 7 years.
  • The Circuitry: The ICD circuitry determines how and when both bradycardia pacing and anti-tachycardia therapies are delivered. This facilitates a wide range of programming options so that each device can be tailored to function in a manner most appropriate for each individual. The ICD capacitor(s) is (are) a critical electrical element allowing for defibrillation. Capacitors perform this task by storing energy (charging) that can then be released across the heart (defibrillation/ cardioversion).

The ICD header and can

The ICD generator communicates with the heart through a ventricular defibrillator ± atrial and LV electrode or lead. The leads are connected to the ICD via a header. The header provides holes for inserting all aspects of the ventricular defibrillator lead and, if needed, an atrial and LV lead in specific ICD models. As in the pacemaker, set screws in the header may be tightened to fix the leads in place or loosened to allow their removal. The metal casing of the ICD generator is called the “can.” The can is similar in composition to the pacemaker's and protects the ICDs' electrical elements from fluids and many external electrical sources.

Ventricular ICD leads

A ventricular defibrillator lead consists of multiple internally separated metal wires externally encased in silicone rubber or polyurethane insulation. This allows the lead to function similarly to a standard pacemaker lead, in other words, the transmission of electrical pacing/sensing signals between the heart and generator, but also structurally provides a separate pathway that participates in the delivery of shocks. This pathway includes what is commonly referred to as the lead “coil(s).” Shocks are delivered across the heart between the coil(s) and potentially the ICD can. Each lead element has its pin that may connect to the ICD header. Ventricular ICD leads commonly have 1 (single) or 2 (dual) coils. Both types have a coil distally, but a dual coil lead also has a proximal one along the length of the lead that preferably winds up in the superior vena cava (SVC) when this type of lead is desired. Thus, each ventricular defibrillator lead has 2 or 3 pins that connect to the ICD, 1 for pacing and sensing and 1 or 2 additional for defibrillating, depending on whether it is a single or dual coil lead. The ventricular ICD lead is essentially a bipolar lead, either an integrated bipolar configuration that utilizes the lead tip and the distal coil for pacing and sensing purposes or a true bipolar configuration that utilizes the tip and a ring electrode.

Complications

Complications are mainly related to ICD implantation and are uncommon but may include:

  • Infection at the implant site that may extend from a mild reaction and seroma up to suppuration and necessitates lead extraction
  • Allergic reaction to medications used during the procedure
  • Bleeding or bruising where your ICD was implanted
  • Damage to the vein where your ICD leads are placed
  • In rare cases, upper limb deep vein thrombosis (DVT)
  • Bleeding around the heart, which can be life-threatening. It may include pericardial effusion and may lead to tapping and surgical intervention
  • Pneumothorax during the procedure of implantation.

Clinical Significance

ICD Functions

Sensing

Sensing in implantable devices refers to the ability to pick up intrinsic signals from the heart and interpret them properly in such a way that allows the device to respond appropriately. The ICD system needs to be sensitive enough to detect the VF wavelength that occurs in a tenth of millivolts and also not to over-sense the T wave, far fields, or myopotential. This is usually achieved through a digitally modified signal algorithm. The peak signal amplitude of the sensed ventricular event during the refractory period is stored to determine the threshold start. When the sensed refractory period expires, the threshold start (which is a percentage of the peak sensed ventricular event, usually 50%) determines the sensitivity setting of the next cycle. Three maximum sensitivity settings are used in the ICD: 2 are used for the pacemaker function of the ICD, and the third is the threshold start for the defibrillation function.

Arrhythmia detection

Proper ICD therapy is a matter of life. Most ICDs have 3 broad detection categories: normal sinus rhythm, ventricular tachycardia, and ventricular fibrillation. These rhythms are defined by rate; the device determines based on the corresponding cycle lengths or millisecond (ms) intervals.

Most ICDs now offer 3-tiered therapy to terminate the tachycardia:

  • Anti-tachycardia pacing (ATP)
  • Cardioversion or low-energy shocks
  • Defibrillation or high-energy shocks.

Detection is usually defined according to the ability of the ICD to detect the rate of the arrhythmia.

  1. The defibrillation-only configuration detects only the sinus rhythm with a rate of less than 200 bpm and VF with a rate of more than 200 bpm. It offers only defibrillation to rates above 200 bpm.
  2. Defibrillation and single tachycardia zone: In this configuration, the ICD can identify 3 detection rates: sinus rhythm with a rate of less than 120 bpm, VT between 120 bpm and 200 bpm, and VF at a rate of more than 200 bpm. It offers ATP for the VT detection zone and defibrillation for the VF detection zone.
  3. Defibrillation with 2 tachycardia zones:
  • The ICD has a sinus rate of less than 120 bpm, slow VT at tachycardia zone A with a rate between 120bpm to 160 bpm, fast VT at 160 bpm to 200 bpm, and a VF zone at a rate of more than 200 bpm. The ICD usually uses the average interval together with the current interval. If the current interval meets the average rate, the interval is binned; if the current rate does not meet the average interval, then the rate is binned to the faster rate. This helps overcome the occasional single-beat rhythm variation, resulting in therapy delivery. 
  • There are 2 methods for counting the sensed events and binning them into the accurate rate zone. The first method is the consecutive interval counting, determined by a programmable number of events needed to satisfy detection, within the tachycardia zone, each interspersed non-tachycardia beat causes counting to be reset to zero. The other method is the X of Y counting method, the continuous rolling detection window updated with each new ventricular heartbeat. Detection is initially satisfied when X out of the total number of beats in the window Y is classified as a tachycardia zone. X of Y scheme is the counting method for VF zone detection. Also, VT can have some variation in its cycle length. An X of Y counting method may compensate for potential transient drops below the rate cutoff and prevent delayed detection and delivery of therapy.

SVT/VT discrimination

In addition to rate detection, rhythm detection is crucial in therapy delivery. Heart rate is not the only factor that detection in VT zones may rely upon. SVT discriminators are algorithms available in the Tach A and Tach B zones. (SVT discriminators cannot be used in the fib zone.)

  • Rate branch: Rate Branch is 1 of the discrimination methods, by simultaneous counting of both atrial and ventricular events, in case of supraventricular events the atrial events are more than ventricular events, then the device diagnoses this as atrial flutter or AF. This is an SVT, and the ICD would automatically inhibit therapy. On the other hand, if V is greater than A, that is, the ventricular rate was more rapid than the atrial rate, the ICD would categorize this as VT and advance to deliver therapy.
  • Interval stability: Interval stability is another method to overcome rapid ventricular rate associated with the chronic Afib patients, tachycardia originating in the ventricles usually has a more regular, albeit rapid rhythm. Thus, interval stability is a special SVT discriminator that assesses the stability of the ventricular rate. Interval stability in the form of regular R–R intervals is diagnosed as a VT (therapy is delivered), while instability in the R–R intervals is diagnosed as AF with rapid ventricular response, inhibiting therapy. The programmable setting for the interval stability of the R-R interval may be a delta or an interval of 80 msec; if the R-R interval changes by more than 80 msec, the diagnosis of Afib is set up, and thus, therapy is inhibited, while if R-R variability is less than the programmable delta, then the diagnosis of VT is established, and therapy is delivered.
  • Sudden onset: This algorithm is used when VT and sinus tachycardia are confused. Sinus tachycardia is characterized by a gradual onset due to physiological demand; however, VT starts abruptly. The sudden onset algorithm looks at interval measurement change (delta) from the non-tachycardia to the binned tachycardia event. VT is diagnosed when the change is more than the programmed value.
  • Morphology discrimination: This algorithm needs a highly expert clinician who uses the own patient's rhythm as a template during programming session to discriminate between supraventricular ORS morphology that uses the His-Purkinje conduction system to match it to the arrhythmia interval detected to make a matching score of the arrhythmic ORS complex.

Therapy

Ventricular tachyarrhythmia therapy in an ICD may be programmed for each detection zone. For the VF zone, therapy occurs in the form of defibrillation. Shocks to terminate tachycardias in this zone may occur regardless of the timing in the cardiac cycle and up to a maximum of 25 J to 42 J stored energy in contemporary devices. Some ICD models may deliver up to 8 shocks per arrhythmia episode.

ATP may end arrhythmia in the VT zone; the ventricular pacing is delivered in a burst at a cycle length shorter than the tachycardia (typically 10% to 20% shorter). The tachycardia may be terminated by rendering its circuit refractory at a critical point. ATP burst options include pacing at a fixed cycle length, a progressively shorter cycle length (“ramp”) with each impulse, and a combination of both. The number of pacing impulses within each burst and the number of burst attempts are also programmable. Cardioversion is the next level for VT termination; each shock is synchronized to occur on R wave in the cardiac cycle, the initial shock energy is usually low and is increased in subsequent shocks.

Bradycardia pacing

ICD systems have bradycardia pacing capabilities similar to those of conventional pacemakers. Whether or not a backup pacing function is required for the defibrillator patient may frequently depend on why the ICD was implanted and the future need for medication that may precipitate significant Bradyarrhythmias.

Enhancing Healthcare Team Outcomes

The cardiologist or cardiac surgeon often performs ICD implantation. However, the primary care provider often cares for these patients. These healthcare professionals should closely monitor the patients for shocks and arrhythmias. Any adverse event should be referred promptly to a cardiologist. In addition, any patient who is a survivor of cardiac arrest, has structural heart disease, Unexplained syncope, Ischemic cardiomyopathy, Non-ischemic dilated cardiomyopathy, LV dysfunction due to prior MI, or has Non-sustained VT should be referred to a cardiologist to determine if they are candidates for an ICD.

References


[1]

Bokeria LA, Neminushchiy NM, Postol AS. Implantable Cardioverter-Defibrillators are the Main Link in the Modern Concept of Sudden Cardiac Death Prevention. Problems and Prospects of the Development of the Method. Kardiologiia. 2018 Dec 25:58(12):76-84. doi: 10.18087/cardio.2018.12.10197. Epub 2018 Dec 25     [PubMed PMID: 30625100]


[2]

Kovacs B, Reek S, Krasniqi N, Eriksson U, Duru F. Extended Use of the Wearable Cardioverter-Defibrillator: Which Patients Are Most Likely to Benefit? Cardiology research and practice. 2018:2018():7373610. doi: 10.1155/2018/7373610. Epub 2018 Nov 29     [PubMed PMID: 30622822]

Level 2 (mid-level) evidence

[3]

Merinopoulos I, Corballis N, Eccleshall SC, Vassiliou VS. Risk of sudden cardiac death: Are coronary chronic total occlusions an additional risk factor? World journal of cardiology. 2018 Dec 26:10(12):250-253. doi: 10.4330/wjc.v10.i12.250. Epub     [PubMed PMID: 30622683]


[4]

Ogano M, Iwasaki YK, Tsuboi I, Kawanaka H, Tajiri M, Takagi H, Tanabe J, Shimizu W. Mid-term feasibility and safety of downgrade procedure from defibrillator to pacemaker with cardiac resynchronization therapy. International journal of cardiology. Heart & vasculature. 2019 Mar:22():78-81. doi: 10.1016/j.ijcha.2018.12.012. Epub 2018 Dec 29     [PubMed PMID: 30619931]

Level 2 (mid-level) evidence

[5]

Orgeron GM, Bhonsale A, Migliore F, James CA, Tichnell C, Murray B, Bertaglia E, Cadrin-Tourigny J, De Franceschi P, Crosson J, Tandri H, Corrado D, Calkins H. Subcutaneous Implantable Cardioverter-Defibrillator in Patients With Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia: A Transatlantic Experience. Journal of the American Heart Association. 2018 Nov 6:7(21):e008782. doi: 10.1161/JAHA.118.008782. Epub     [PubMed PMID: 30608223]


[6]

Ahmed FZ, Fullwood C, Zaman M, Qamruddin A, Cunnington C, Mamas MA, Sandoe J, Motwani M, Zaidi A. Cardiac implantable electronic device (CIED) infections are expensive and associated with prolonged hospitalisation: UK Retrospective Observational Study. PloS one. 2019:14(1):e0206611. doi: 10.1371/journal.pone.0206611. Epub 2019 Jan 2     [PubMed PMID: 30601808]

Level 2 (mid-level) evidence

[7]

Osmanska J, Jhund PS. Contemporary Management of Heart Failure in the Elderly. Drugs & aging. 2019 Feb:36(2):137-146. doi: 10.1007/s40266-018-0625-4. Epub     [PubMed PMID: 30535931]


[8]

Bernier R, Al-Shehri M, Raj SR, Reyes L, Lockwood E, Gulamhusein S, Williams R, Valtuille L, Sivakumaran S, Hruczkowski T, Kimber S, Exner DV, Sandhu RK. A Population-Based Study of Adherence to Guideline Recommendations and Appropriate-Use Criteria for Implantable Cardioverter Defibrillators. The Canadian journal of cardiology. 2018 Dec:34(12):1677-1681. doi: 10.1016/j.cjca.2018.08.030. Epub 2018 Aug 24     [PubMed PMID: 30527158]


[9]

Kazemi-Arpanahi H, Vasheghani-Farahani A, Baradaran A, Mohammadzadeh N, Ghazisaeedi M. Developing a Minimum Data Set (MDS) for Cardiac Electronic Implantable Devices Implantation. Acta informatica medica : AIM : journal of the Society for Medical Informatics of Bosnia & Herzegovina : casopis Drustva za medicinsku informatiku BiH. 2018 Oct:26(3):164-168. doi: 10.5455/aim.2018.26.164-168. Epub     [PubMed PMID: 30515006]


[10]

Aslian H, Delana A, Kaiser SR, Moretti E, Foti C, Bregant P, de Denaro M, Longo F, Severgnini M. A multicenter dosimetry study to evaluate the imaging dose from Elekta XVI and Varian OBI kV-CBCT systems to cardiovascular implantable electronic devices (CIEDs). Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2018 Nov:55():40-46. doi: 10.1016/j.ejmp.2018.10.013. Epub 2018 Oct 22     [PubMed PMID: 30471818]